Entrance to departments ie wards, theatres, etc or at nurse stations.Connected to the medical gas pipeline system via pressure sensors mounted directly to the pipeline. Signals are transmitted to the medical gas alarm situated in areas of high visibility for nursing staff.
BeaconMed?s Digital Medipoint 26 MK2 Medical Gas Local Area Alarm is designed to monitor up to 6 gas services in the normal, high and low pressure conditions. They are suitable for surface or flush wall installation or integration into Area Valve Service Modules (Medizones). Repeater panels can be easily connected to provide warnings in other areas for example one panel outside the Operating Theatre and one panel inside. Pressure sensors for alarms are installed downstream of the area valve service units to provide local warnings of supply pressure/vacuum outside of the normal operating range. Line contact monitoring circuits constantly monitor the integrity of the pressure sensors and interconnecting wiring for added safety.
For more information on Medical Gas Area Alarm Digital Medipoint 26 talk to BeaconMedaes