Quick and simple to use Tests all equipment 110V, 230V and 415V
Tests IT equipment (computers, etc.) Earth Bond Test at 25A
Limits on sub leakage up to 0.75mA Built in buzzer and timer for all tests
Individual pass and fail of all tests by LED display Weighs less than 3.5kg
Overall pass and fail of equipment indicated by LED display
Earth Bond pass levels indicated better than 0.1ž or 0.5ž
All tests timed and carried out to meet the NEW IEE Guidance notes and EAWR 1989 plus HSE requirements
Complete with rugged carrying bag for easy operation, and calibration certificate
Meets the requirements of the NEW IEE Guidance notes on pass levels for insulation test
i.e. 1Mohm for Class 1 Equipment & 2Mohm for Class 2 Equipment