Sign In
The Ametek Jofra ETC-125A is a simple yet effective tool for verifying temperature instruments that also require references below ambient temperatures: e.g. air-conditioning and cold counters. The predrilled holes allow the use of an insertion tube in the largest bore. This increases the flexibility to match many sensor-under-test sizes.
All instrument controls are accessed directly from the front panel. The main functions on the ETC series are designed with one-key-one-function logic. This means that there are no difficult multiple keystrokes to remember to access primary functions. The easy-to-read, backlit display features dedicated icons, which help in identifying instrument conditions and operational steps.
The Jofra ETC-125A has a temperature range of -10 to 125°C (14 to 257°F). The small size and light weight make it a perfect instrument to verify sensors in difficult to reach places.
ETC SERIES FEATURES
Temperature range from -10 to 400°C (14 to 752°F) with 3 models.
Accuracy to + 0.5°C and Stability to + 0.05°C.
Time saving fast cooling and heating times.
Easy-to-read multi-information display.
Customizable stability checkmark with countdown timer.
Set exact temperature desired to a resolution of 0.1°C.
ETC 400R, small lightweight IR calibrator.
Lightweight and easy to carry.
Ametek Jofra ETC-125A Specifications
Temperature range @ ambient temp. 23°C / 73°F
Maximum
125°C / 257°F.
Minimum @ ambient temp. 0°C / 32°F
Minimum @ ambient temp. 23°C / 73°F
Minimum @ ambient temp. 40°C / 104°F
-18°C / -0°F
-10°C /14°F
6°C / 43°F.
Resolution
Selectable
1° or 0.1°.
Heating Time
-10 to 23°C / 14 to 73°F
23 to 100°C / 73 to 212°F
100 to 125°C / 212 to 257°F
3 minutes
11 minutes
7 minutes.
Cooling Time
125 to 100°C / 257 to 212°F
100 to 0°C / 212 to 32°F
0 to -10°C / 32 to 14°F
1 minutes
17 minutes
14 minutes.
Stability
+0.05°C / +0.09°F.
Measured after the stability indicator has been on for 10 minutes. Measuring time is 30 minutes.
Time to stability
3 minutes (approximate).
Accuracy
+0.5°C / +0.9°F.
General Specifications
Immersion depth
110 mm / 4.3 in (insulation included).
Voltage AC
115V (90-132) and 230V (180-264).
Frequency
47 - 63 Hz.
Power consumption (max.)
75 VA
Serial data interface
RS232.
Operating temperature
0 to 40°C / 32 to 104°F.
Storage temperature
-20 to 50oC / -4 to 122°F.
Humidity
0 to 90% RH.
Protection class
IP10.
DNV Marine Approval, Certificate no
A-10384.
Dimensions (L x W x H)
172 x 72 x 182 mm / 6.8 x 2.8 x 7.2 in.
Weight
1.8 kg / 3.9 lb.
Ametek ETC-125A warranty
Two-years.