Sign In
An easy-to-use, high-performance instrument, the Chauvin C.A 6255 10A micro-ohmmeter tests the quality of connections with precision (resolution of 0.1 ?Ω) through the use of low resistance measurements. Housed in a rugged construction-site type case with cover, the C.A 6250 is a self-contained instrument powered by a rechargeable battery with built-in charger.
It provides 7 measurement ranges, from 5 mΩ to 2,500 Ω, that can be accessed and selected directly on the front-panel rotary switch. It uses the 4-wire measurement method, with automatic compensation of spurious voltages.
This device is compliant with safety standard IEC 61010-2-030 and the leads are compliant with IEC 61010-031, for voltages up to 50 V with respect to earth in category III.
3 Measurement Modes
NON-INDUCTIVE MODE - for measurements on resistors with a time constant of a few ms (contacts, metallisation, etc.).
INDUCTIVE MODE - for measurements on inductive components (transformers, motors, etc.).
NON-INDUCTIVE MODE - with AUTOMATIC trigger: for measurements on resistors without a time constant: for repetitive measurements in production, helping to gain considerable time.
Temperature Compensation
The value of a resistance varies according to the temperature. For reliable supervision allowing the comparison of measurements which are comparable because they were made in the same conditions, it is a good idea always to express a measurement in terms of a given reference temperature. You can perform this calculation by simply pressing a key.
FEATURES
Automatic detection of the presence of an external AC or DC voltage on the terminals
3 different measurement modes depending on the nature of the resistance to be measured,
Protection of the operator when a resistance having a large inductive component (motor, transformer, etc.) is measured.
Measurement up to 2,500 Ω, resolution 0.1 µΩ.
Programming of alarm thresholds.
Measurement temperature using a Pt100 jack on the front panel,
Function for automatic calculation of the resistance at a reference temperature
Extended memory making it possible to store approximately 1,500 measurements,
Indication of the level of memory use.
Indication of battery charge condition.
Automatic switching of the backlighting to standby to save battery power,
RS 232 interface to print the results on a serial printer or export them to a PC.
Chauvin C.A 6255 Specifications
4-wire measurement with compensation of spurious voltages.
RANGE
RESOLUTION
ACCURACY OVER 1 YEAR
MEASURING CURRENT
VOLTAGE DROP
5.000 mΩ
0.1 ?Ω
0.05% + 1.0 ?Ω
10 A
50 mV
25.000 mΩ
1 ?Ω
0.05% + 3 ?Ω
10 A
250 mV
250.00 mΩ
10 ?Ω
0.05% + 30 ?Ω
10 A
2500 mV
2500.0 mΩ
0.1 mΩ
0.05% + 0.3 mΩ
1 A
2500 mV
25.000 Ω
1 mΩ
0.05% + 3 mΩ
100 mA
2500 mV
250.00 Ω
10 mΩ
0.05% + 30 mΩ
10 mA
2500 mV
2500.0 Ω
100 mΩ
0.05% + 300 mΩ
1 mA
2500 mV
Measurement modes
Inductive, non-inductive, non-inductive with automatic trigger.
Temperature compensation
By temperature sensor or manual.
Data storage
1,500 measurements.
Operating temperature
0°C to +50°C.
Electrical safety
EN 61010-1.
Max. voltage relative to earth:
50 V.
Mechanical protections
IP53 = housing open.
IP64 = housing closed.
Dimensions
270 x 250 x 180mm.
Weight approx.
4 kg.
Chauvin C.A6255 warranty
Two years.
For more information on Chauvin C.A 6255 Microhmmeter talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.