The Chauvin C.A 6505 megohmmeter is a portable instrument housed in a rugged field case with lid and operates on either battery or line power. It makes voltage, insulation, and capacitance measurements.
As soon as the switch is set to an insulation measurement position, the instrument automatically measures the presence of any AC/DC voltage. This voltage is measured at all times and indicated on the small display unit. The CA6505 automatically determines AC or DC: the AC measurement is an RMS value.
Insulation is sensitive to temperature and humidity variations. In addition, measurement is affected by the appearance of disturbance currents. To overcome these effects, long-term measurements must be performed and the PI and DAR coefficients must be calculated. They will then allow qualification of the insulants and their ageing.
FEATURES
Automatic voltage measurement,
Automatic detection of the presence of an external AC or DC voltage on the terminals.
Simple user interface,
Calculation of the PI and of the DAR.
Fuse protection of the instrument, with detection of a defective fuse.
Operator safety thanks to the automatic discharging of the device tested.
Automatic power-down of the instrument to extend battery life.
Indication of the battery charge condition.
Large backlit LCD display unit with many announciators for very easy reading.
Chauvin C.A 6505 Specifications
Fixed test voltages
500 V
1000 V
2500 V
5000 V
Insulation ranges
10 kΩ to 2 TΩ
10 kΩ to 4 TΩ
10 kΩ to 10 TΩ
10 kΩ to 10 TΩ
Adjustable test voltage
40 V to 5,100 V in 10 or 100 V increments.
Voltage
1 to 5,100 V (15 Hz to 500 Hz or DC).
Capacitance
0.001 to 49.99 F.
Leakage current measurement
0.00 - 3mA.
Functions
Display
Large LCD screen.
Backlighting
Programmable alarms
No.
Smoothing of display
No.
Prog. test duration
Calculation of ratios
DAR - PI.
Step voltage mode
Calc. of R at reference T°
No.
Locking of test voltage
R(t)
No.
Storage
RS 232
Printing of measurements
PC software
Power supply
Rechargeable NiMH battery.
Battery life
30 days with 10 DARs and 5 PIs / day.
Electrical safety
IEC 61010-1 (Cat. III 1,000 V or Cat. I 2,500 V) and 61557.
Dimensions (L x W x H)
270 x 250 x 180 mm.
Weight
4.3 kgs.
Chauvin C.A6505 warranty
12 months.
For more information on Chauvin C.A 6505 talk to Cuthbertson Laird Group
Enquire Now