Sign In
The Druck LiquidTC 165 combines the portability of a dry block calibrator with the flexibility of a liquid immersion bath to enable the testing and calibration of virtually any shape and size of devices.
The calibrators are factory configured as liquid baths and are provided with a bath cover to hold up to 5 devices while reducing heat loss from the surface of the liquid medium.
For transportation a leak-proof sealing cover is also provided as standard. Optionally the temperature calibrators can be configured with additional capabilities including interchangeable liquid baths, a black body source and dry block interchangeable inserts.
FEATURES
LiquidTC165 Liquid bath temperature calibrator -35°C to 165°C.
Standard Liquid Bath Configuration.
Accuracy from 0.1°C.
Stability 0.05°C.
Large bath for irregular and multiple devices.
Multi-purpose - liquid bath, black bodysource, dry block.
Interchangeable bath - simplifies fluidchanges.
Light weight and robust for field use.
Leak-proof bath cover for transportation.
Druck LiquidTC 165 Specifications
Temperature Range
Room Temperature (RT) to 155°C (-30° to 310°F option) with TCL 10 oil supplied
25° to 165°C (-77° to 330 °F option) with optional TCL 50 oil.
Heat Well
Dia 60 mm/depth 170 mm (150 mm working depth).
Measuring Zone
0 to 40 mm from the bottom of the insert.
Stability
0.05°C.
Accuracy
Standard Liquid Bath - LI
0.2°C.
Dry Block Option - DB
0.3°C.
Infrared Black Body Option - IR
0.5°C.
Black Body Emissivity
0.9994.
Display
Display Range
-50 to 165°C.
Measurement Resolution
0.01° from -9.99 to 99.99 otherwise 0.1°C.
Setting Resolution
0.1°C.
Standard Features
Functions
Controller off, manual control, operating service hours, set-point memory, rate control °C/min, test profile.
Power Requirements
Supply Voltage
100 to 240 V AC 50/60 Hz.
Power Consumption
400 VA nominal.
Dimensions/Weight
Width
210 mm.
Height
380 + 50 mm.
Depth
300 mm.
Weight
12.5 kg.
Environmental
Ambient Temperature
10°C to 50°C (50°F to 122°F)
Specifications valid at 18 °C to 28 °C (65 °F to 82 °F) ambient temperature. Minimum and maximum temperatures may not be achieved over the full ambient temperature range.
Ambient Humidity
To 80% RH (non-condensing).
Storage Temperature
From -10 to 60°C (14 to 140°F).
Recommended Operating Environment
Indoor use. Not for use in potentially explosive atmospheres.
Compliance
EN6326 (EMC), EN61010 Oelectrical safety), EU directive (RoHS, REACH, WEEE).
Druck LiquidTC 165 warranty
One-year.
For more information on Druck LiquidTC 165 Liquid Bath Temperature Calibrator talk to Cuthbertson Laird Group
Enquire Now