Sign In
One of Fluke best-selling products is the Tweener thermometer, and there’s a reason. No other company, not one, has a thermometer that comes close to the performance and features of the Tweener. The 1502A Tweener features accuracy up to + 0.006°C.
Each Tweener is programmable to match a probe’s constants for maximum linearity and accuracy. All probe constants and coefficients are programmed through simple, front-panel keystrokes. Temperature is displayed in °C, °F, K, or resistance in ohms.
The 1502A is calibrated digitally using the front-panel buttons. You never have to open the box to calibrate it. This calibration protocol further reduces the cost of the 1502A. It goes where you go and works the way you want it to.
FEATURES
Measures 25Ω and 100Ω RTDs and SPRTs.
Four-wire connection eliminates lead resistance effects.
Fast one-second measurement cycle.
Adjustable digital filter.
Accepts ITS-90 characterization coefficients.
Also accepts Callendar-Van Dusen and IPTS-68 coefficients.
Adjustable excitation current.
Password protection of critical parameters.
Serial RS-232 interface standard; IEEE-488 GPIB interface optional.
Detachable power cord.
Light weight, small and portable.
Sturdy, reliable construction.
Fluke 1502A Specifications
Temperature Range?
?200 °C to 962 °C (?328 °F to 1764 °F).
Resistance Range
0 Ω to 400 Ω, auto-ranging.
Probe
Nominal RTPW: 10Ω to 100Ω
RTD, PRT, or SPRT.
Characterisations
ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11
IPTS-68: R0, a, d, a4, and c4
Callendar-Van Dusen: R0, a, d, and b.
Resistance Accuracy (ppm of reading)
0Ω to 20Ω: 0.0005W
20Ω to 400Ω: 25 ppm.
Temperature Accuracy?
+0.004 °C at ?100 °C
+0.006 °C at 0 °C
+0.009 °C at 100 °C
+0.012 °C at 200 °C
+0.018 °C at 400 °C
+0.024 °C at 600 °C.
Operating Temperature Range
16 °C to 30 °C.
Resistance Resolution
0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω.
Temperature Resolution
0.001 °C.
Excitation Current
0.5 and 1 mA, user selectable, 2 Hz.
Measurement Period
1 second.
Digital Filter
Exponential, 0 to 60 seconds time constant (user selectable).
Probe Connection
4-wire with shield, 5-pin DIN connector.
Communications
RS-232 serial standard
IEEE-488 (GPIB) optional.
Display
8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters.
Power
115 VAC (+10 %), 50/60 Hz, 1 A, nominal
230 VAC (+10 %), 50/60 Hz, 1 A, nominal, specify.
Probes from Hart
5615, 5627, 5626, 5628, 5622.
Calibration
ISO 17025-accredited calibration provided.
Dimensions (H x W x D)
61 x 143 x 181mm (2.4 x 5.6 x 7.1 in).
Weight
1.0 kg (2.2 lb.).
Fluke 1502A 'Tweener' Warranty
1 year.
?Temperature ranges and accuracy may be limited by the sensor you use.
For more information on Fluke 1502A 'Tweener' PRT Thermometer talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.