Sign In
The Fluke 1552A “Stik” Thermometer is accurate, stable and durable, making it the best alternative to mercury thermometers and existing electronic thermometers. The thin-film RTD sensor incorporates the same characteristics of other premium reference thermometers designed by Fluke calibration, but it is more robust and less susceptible to drift.
The stainless steel probe sheath of the “Stik” Thermometer is far more robust than the glass stem of a mercury thermometer and better suited for industrial environments. And since the “Stik” thermometer does not contain mercury, the risk of an accidental mercury spill is eliminated.
The digital display and probe are fixed together as a system. The hinge allows the display to rotate 90° to easily view the measurement from any angle.
FEATURES
Accuracy of + 0.05 °C (+ 0.09 °F) over full range.
Intrinsically safe (ATEX and IECEx compliant).
User-configurable temperature Trend/Stability indicator.
Display temperature in °C or °F.
Optional data logging to internal memory.
300-hour battery life.
Percent battery-life and low-battery indicator.
NVLAP-accredited, NIST-traceable calibration included.
Fluke 1552A Specifications
Temperature Range
-80 °C to 300 °C (-112 °F to 572 °F).
Accuracy (1 year)
+ 0.05 °C (+ 0.09 °F).
Display units
°C, °F.
Sensor type
10Ω wire-wound PRT.
Probe temperature coefficient
0.00385 Ω/Ω/ΩC nominal.
Sensor length
? 30 mm.
Sensor position (from sheath tip)
3 mm (0.1 in).
Minimum immersion depth
12 cm (4.8 in).
Probe sheath material
Stainless steel.
Response time
4.8 mm diameter probe (3/16 in): 14 seconds
6.35 mm diameter probe (1/4 in): 21 seconds.
Probe hysteresis
+ 0.01 °C.
Temperature resolution
Selectable: 0.1, 0.01, 0.001 (default 0.01).
Sample rate
Selectable: 0.5 seconds, 1 seconds, 2 seconds (default 1 seconds).
Operating temperature range readout
-10 °C to 50 °C (14 °F to 122 °F).
Storage temperature
-20 °C to 60 °C (-4 °F to 140 °F) 0 % to 95 % RH, non-condensing.
Optional data logging1
Up to 10,000 time-stamped readings stored to internal memory.
Logging intervals1
2 seconds, 5 seconds, 10 seconds, 30 seconds, or 60 seconds;
2 minutes, 5 minutes, 10 minutes, 30 minutes, or 60 minutes
Damping
Moving average of the most recent 2 to 10 readings (ON/OFF, 2, 5, 10).
Communications
RS-232 stereo jack (access calibration parameters only).
AC power
None.
DC Power
3-AAA Batteries, typical battery life of 300 hours without LCD backlight.
EMC compliance
EN61326:2006 Annex C; CISPR II Edition 5.0-2009; Class “B.
Enclosure rating
IP50.
Electronics dimensions (H x W x D)
114 mm x 57 mm x 25 mm (4.5 in x 2.25 in x 1.0 in).
Weight
196 g (6.9 oz).
Calibration (included)
NVLAP-accredited, NIST-traceable.
Characterisation
ITS-90.
Fluke 1552A warranty
1 year.
1. For Data-Logging option, please choose the 1552A-12-DL option from the dropdown.
For more information on Fluke 1552A-12 Intrinsically Safe 'Stik' Thermometer talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.