Sign In
*** Please Note: When ordering also specify the termination you require, ie. Fluke 5618B-6-G (gold pins). X = termination ***
Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).
All models come without calibration unless calibration is purchased separately.
Featuring a 3.2 mm diameter (1/8 in) sheath, these industrial standards probes have reduced response time without compromising precision. This small diameter 5618A probe works well in many applications where immersion depth is limited. Larger diameter probes give more measurement error in short immersion depth applications because they conduct more heat between ambient and the sensor.
The Fluke 5618B is also a great probe to use for calibrating your Fluke Calibration 9132 or 9133 infrared calibrators. In fact, these IR black body heat sources were designed to be calibrated with this type of probe. We recommend using the Fluke 5618B PRTs with the 1523, 1524, 1502A, 1529, or 1560 thermometer readouts.
FEATURES:
152 mm (6 in) Small Diameter Probe, lead wire (cable) 6 ft.
Small diameter sheath, 3.2mm (0.125 in).
Excellent stability.
Includes ITS-90 coefficients.
NVLAP-accredited calibrated from -200°C to 500°C, lab code 200706-0.
Fluke 5618-6 Specifications
Resistance
Nominal 100 Ω at 0 °C.
Temperature
Coefficient
0.003923 Ω/Ω/ °C nominal.
Temperature
Range
-200 °C to 500 °C
(-200 °C to 300 °C for 5618B-6-X).
Drift Rate
+ 0.1 °C when used periodically to 500 °C.
Sheath Material
316 SST.
Leads
22 AWG PTFE, 6’.
Termination
Specify.
Hysteresis
Less than 0.01 °C at 0 °C when using -196 °C and 420 °C as the end points.
Time Constant
Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
Thermal EMF
Less than 25 mV at 420 °C.
Calibration
Includes manufacturer’s NVLAP-accredited calibration w/ITS-90 coefficients, R vs. T values in 1 °C increments, lab code 200706-0.
Size
5618B-12: 305 mm L x 3.2 mm dia. (12 x 1/8 in)
5618B-9: 229 mm L x 3.2 mm dia. (9 x 1/8 in)
5618B-6: 152 mm L x 3.2 mm dia. (6 x 1/8 in)
Calibrated
Accuracy? (k=2)
+ 0.026 °C at -200 °C
+ 0.066 °C at 0 °C
+ 0.219 °C at 500 °C.
?Includes
calibration
uncertainty and
drift specification.
NVLAP? Calibration Uncertainty
Temperature
Expanded
Uncertainty
(k=2)
-196 °C
0.024 °C
-38 °C
0.011 °C
0 °C
0.010 °C
200 °C
0.018 °C
420 °C?
0.029 °C
Note: Calibration uncertainties depend on the uncertainties of the lab performing the calibration. Subsequent calibrations of this probe performed with different processes, at different facilities, or with changed uncertainty statements may state different uncertainties. ?Lab code 200706-0 ?300 °C for 5618B-6.
For more information on Fluke 5618B-6-X Small Diameter Industrial RTD talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.