Sign In
The Fluke 7103 Micro-Bath may be used as a portable instrument or bench top temperature calibrator for calibrating thermocouple and RTD temperature probes. With an ambient temperature of 23°C (74°F), calibrations may be done over a range of -5°C to 126°C (23°F to 257°F). The resolution of the Fluke 7103 temperature display is 0.01°.
All Micro-Baths have RS-232 ports, come with our Interface-it software, and can be used with MET/TEMP II software. Also included are contacts to calibrate a thermal switch, eight set-point memory storage, ramp-rate adjust, and over-temperature safety cutout.
FEATURES
World’s smallest portable calibration baths.
Calibrates sensors of any size or shape.
Stability to +0.03 °C.
Range from -30°C to 125°C.
Fluke 6102, 7102 & 7103 Specifications
Specifications
6102
7102
7103
Range
35°C to 200°C (95°F to 392°F)
-5°C to 125°C (23°F to 257°F)
-30°C to 125°C (-22°F to 257°F)
Accuracy
+0.25°C
+0.25°C
+0.25°C
Stability
+0.02°C at 100°C (oil 5013)
+0.03°C at 200°C (oil 5013)
+0.015°C at -5°C (oil 5010)
+0.03°C at 121°C (oil 5010)
+0.03°C at -25°C (oil 5010)
+0.05°C at 125°C (oil 5010)
Uniformity
+0.02°C
+0.02°C
+0.02°C
Resolution
0.01°C/°F
0.01°C/°F
0.01°C/°F
Operating temperature
5°C to 45°C
5°C to 45°C
5°C to 45°C
Heating time
25°C to 200°C: 40 minutes
25°C to 100°C: 30 minutes
25°C to 100°C: 35 minutes
Cooling time
200°C to 100°C: 35 minutes
25°C to 0°C: 30 minutes
25°C to -25°C: 45 minutes
Well size
64 mm dia. x 140 mm deep (2.5 x 5.5 in) (working area is 48 mm [1.9 in] in diameter).
Size (W x H x D)
14 x 26 x 20 cm
(5.5 x 10.38 x 8 in)
18 x 31 x 24 cm
(7.2 x 12 x 9.5 in)
23 x 34 x 26 cm
(9 x 13.2 x 10.5 in)
Weight
4.5 kg (10 lb) with fluid
6.8 kg (15 lb) with fluid
9.8 kg (22 lb) with fluid
Volume
0.75 L (1.6 pints)
0.75 L (1.6 pints)
1.0 L (2.11 pints)
Power
115 V AC (+10%), 2.3 A or 230 V AC (+10%), 1.1 A, switchable, 50/60 Hz, 270 W
115 V AC (+10%), 1.8 A or 230 V AC (+10%), 0.9 A, switchable, 50/60 Hz, 200 W
94-234 V AC (+10%), 50/60 Hz, 400 W
Computer interface
RS-232 included with free Interface-it software.
NIST-traceable calibration
Data at 50°C, 100°C, 150°C, and 200°C
Data at -5°C, 25°C, 55°C, 90°C, and 121°C
Data at -25°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 125°C
For more information on Fluke 7103 Micro-Bath Thermometer Calibrator talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.