Sign In
The Fluke 8846A bench multimeter can be incorporated easily into your automated test system. The meter emulates several legacy bench DMMs, allowing them to be substituted easily into existing systems. SCPI compatibility and drivers allow them to work with most available application and development languages.
Front and rear inputs let you easily make connections where it’s most convenient, whether the meter is rack-mounted or used on a bench. Multiple interfaces provide compatibility to existing and new standards.
You can also use the 8846A to measure temperature, capacitance, period, and frequency-the functions of a counter, capacitance meter, and thermometer are built in for unparalleled versatility. Extend the meters’ utility even more with their graphical display modes, including TrendplotTM paperless recorder mode, statistics and histograms - features you won’t find on other multimeters.
The Fluke 8846A ships with a free copy of FlukeView Basic. Transfer data points from your meter to your PC and use the two standard, non-customisable forms to display your readings in table or graphical form.
FEATURES
6.5 digit resolution.
Basic V dc accuracy of up to 0.0024%.
Dual display.
100 µA to 10 A current range, with up to 100 pA resolution.
Wide ohms range from 10 Ω to 1 GΩ with up to 10 µΩ resolution.
2 x 4 ohms 4-wire measurement technique.
Measures frequency and period.
USB memory drive port.
8846A also measures capacitance and temperature.
USB memory drive port (8846A).
Fluke 45 and Agilent 34401A emulation.
Graphical display.
Trendplot™ paperless recorder mode, statistics, histogram.
CAT I 1000 V, CAT II 600 V.
Fluke 8846A Specifications
Display
VFD Dot Matrix.
Resolution
6.5 Digits.
Voltage
V DC Ranges
100 mV to 1000 V.
Max. Resolution
100 nV.
Accuracy
8845A: 0.0035 + 0.0005 (% measurement + % of range)
8846A: 0.0024 + 0.0005 (% measurement + % of range).
V AC Ranges
8845A: 100 mV to 750 V 8846A: 100 mV to 1000 V.
Max. Resolution
8845A: 0.06 + 0.03 (% measurement + % of range)
8846A: 0.06 + 0.03 (% measurement + % of range).
Accuracy
8845A: 0.0035 + 0.0005 (% measurement + % of range)
8846A: 0.0024 + 0.0005 (% measurement + % of range).
Frequency
3 Hz to 300 KHz.
Resistance
2 x 4 Wire
Ranges
8845A: 100 Ω to 100 MΩ
8846A: 10 Ω to 1 GΩ.
Max. Resolution
8845A: 100 µΩ
8846A: 10 µΩ.
Accuracy
8845A: 0.010 + 0.001 (% measurement + % of range)
8846A: 0.010 + 0.001 (% measurement + % of range).
Current
A DC Ranges
100 µA to 10 A.
Max. Resolution
100 pA.
Accuracy
0.050 + 0.005 (% measurement + % of range).
A AC Ranges
8845A: 10 mA to 10 A
8846A: 100 mA to 10 A.
Max. Resolution
8845A: 10 nA
8846A: 100 pA.
Accuracy
0.10 + 0.04 (% measurement + % of range).
Frequency
3 Hz to 10 kHz.
Freq/Period
Ranges
8845A: 3 Hz to 300 kHz
8846A: 3 Hz to 1 MHz.
Max. Resolution
1 µHz.
Accuracy
0.01%.
Continuity/Diode Test
Capacitance
Ranges
8846A: 1 nF to 0.1 F.
Max. Resolution
8846A: 1 pf.
Accuracy
8846A: 1%.
Temperature
Type
8846A: Platinum RTD.
Ranges
8846A: -200 °C to +600 °.
Max. Resolution
8846A: 0.01 °.
Accuracy
8846A: 0.06 °.
Maths Functions
Functions
NULL, Min/Max/Average, Std Dev.
dB/dBm
Advanced Functions
Statistics/Histogram
Trendplot
Limit Test
Input Output
USB Memory
8846A: USB memory drive port.
Real Time Clock
8846A: Yes.
Interfaces
RS 232, IEE-488.2, Ethernet, USB (with optional adaptor).
Programming Languages/Modes
8846A: SCPI (IEEE-488.2), Agilent 34401A, Fluke 45.
General Specifications
Safety
Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V.
Calibration Documentation
17025 accredited report of calibration included.
Dimensions
88 mm x 217 mm x 297 mm (3.46 in x 8.56 in x 11.7 in).
Weight
3.6 kg (8.0 lbs).
Fluke 8846A warranty
Three years.
For more information on Fluke 8846A 6.5 Digit Precision Multimeter talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.