NB. 'X' in the model number to be replaced with A, B, C, D, E, or F as appropriate for the desired insert. See the inserts illustration and listing on the datasheet.
The Fluke 9143 Field Metrology Well is designed to be a reliable, stable heat source that can be used in the field or laboratory. It offers accuracy, portability, and speed for nearly every field calibration application. The instrument has been designed with the field user in mind and is easy to use while maintaining stability, uniformity, and accuracy comparable to some laboratory instruments.
The Fluke 9143-X Field Metrology Well extends high performance to the industrial process environment by maximising portability, speed, and functionality with little compromise to metrology performance. The process version provides a built-in two-channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20 mA transmitters which includes the 24 V loop supply to power the transmitter.
Each process version accepts an ITS-90 reference PRT. The built-in readout accuracy ranges from + 0.01 °C to + 0.07 °C depending on the measured temperature. Reference PRTs for Field Metrology Wells contain individual calibration constants that reside in a memory chip located inside the sensor housing, so sensors may be used interchangeably. The second channel is user-selectable for 2-, 3-, or 4-wire RTDs, thermocouples, or 4-20 mA transmitters.
The process option makes transmitter loop calibrations easy. The transmitter sensor is placed in the well with the reference PRT and the transmitter electronics are connected to the front panel of the instrument. With 24 V loop power, you are able to power and measure the transmitter current while sourcing and measuring temperature in the Field Metrology Well. This allows for the measurement of as-found and as-left data in one self-contained calibration tool.
EASY TO USE
Large LCD display is easy to see even in poorly lighted environments.
Function keys and easy-to-navigate menus.
Visual and audible stability indicator lets you know when you are ready to calibrate.
Pre-programmed calibration routines take the guesswork out of what to do.
One-button push to set the temperature.
Reference readout accepts “plug-and-play”.
smart connector.
Fluke 9143 Product Video
Fluke 9143 Specifications
Temperature Range at 23 °C
33 °C to 350 °C (91 °F to 662 °F).
Display Accuracy
+ 0.2 °C Full Range.
Stability
+ 0.02 °C at 33 °C
+ 0.02 °C at 200 °C
+ 0.03 °C at 350 °C.
Axial Uniformity at 40 mm (1.6 in)
+ 0.04 °C at 33 °C
+ 0.1 °C at 200 °C
+ 0.2 °C at 350 °C.
Radial Uniformity
+ 0.01 °C at 33 °C
+ 0.015 °C at 200 °C
+ 0.02 °C at 350 °C.
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes)
+ 0.015 °C Full Range.
Hysteresis
0.03.
Operating Conditions
0 °C to 50 °C, 0 % to 90 % RH (non-condensing).
Immersion (Well) Depth
150 mm (5.9 in).
Insert OD
25.3 mm (1.00 in).
Heating Time
5 min: 33 °C to 350 °C.
Cooling Time
32 min: 350 °C to 33 °C
14 min: 350 °C to 100 °C.
Resolution
0.01 °.
Display
LCD, °C or °F user-selectable.
Calibration
NVLAP accredited calibration included.
Weight
7.3 kg (16 lb).
Dimensions (H x W x D)
290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in).
Fluke 9143 warranty
1 year.
-P Specifications
Built-in Reference Thermometer Readout Accuracy (4-Wire Reference Probe)?
+ 0.010 °C at -25 °C
+ 0.015 °C at 0 °C
+ 0.020 °C at 50 °C
+ 0.025 °C at 150 °C
+ 0.030 °C at 200 °C
+ 0.040 °C at 350 °C
+ 0.050 °C at 420 °C
+ 0.070 °C at 660 °C.
Reference Resistance Range
0 ohms to 400 ohms.
Reference Resistance Accuracy?
0 ohms to 42 ohms: +0.0025 ohms 42 ohms to
400 ohms: +60 ppm of reading.
Reference Characterisations
ITS-90, CVD, IEC-751, Resistance.
Reference Measurement Capability
4-wire.
Reference Probe Connection
6-pin Din with Infocon Technology.
Built-in RTD Thermometer Readout Accuracy
NI-120: + 0.015 °C at 0 °C
PT-100 (385): + 0.02 °C at 0 °C
PT-100 (3926): + 0.02 °C at 0 °C
PT-100 (JIS): + 0.02 °C at 0 °C.
RTD Resistance Range
0 ohms to 400 ohms.
RTD Resistance Accuracy?
0 ohms to 25 ohms: +0.002 ohms
25 ohms to 400 ohms: +80 ppm of reading.
RTD Characterisations
PT-100 (385),(JIS),(3926), NI-120, Resistance.
RTD Measurement Capability
4-wire RTD (2-,3-wire RTD w Jumpers only).
RTD Connection
4 terminal input.
Built-in TC Thermometer Readout Accuracy
Type J: + 0.7 °C at 660 °C
Type K: + 0.8 °C at 660 °C
Type T: + 0.8 °C at 400 °C
Type E: + 0.7 °C at 660 °C
Type R: + 1.4 °C at 660 °C
Type S: + 1.5 °C at 660 °C
Type M: + 1.4 °C at 660 °C
Type L: + 0.7 °C at 660 °C
Type U: + 0.75 °C at 600 °C
Type N: + 0.9 °C at 660 °C
Type C: + 1.1 °C at 660 °C.
TC Millivolt Range
-10 mV to 100 mV.
Voltage Accuracy
0.025% of reading + 0.01 mV.
Internal Cold Junction Compensation Accuracy
+ 0.35 °C (ambient of 13 °C to 33 °C).
TC Connection
Small connectors.
Built-in mA Readout Accuracy
0.02 % of reading + 2 mV.
mA Range
Cal 4-22 mA, Spec 4-24 mA.
mA Connection
2 terminal input.
Loop Power Function
24 V DC loop power.
Built-in Electronics Temperature Coefficient (0 °C to 13 °C, 33 °C to 50 °C)
+ 0.005 % of range per °C.
Calibration
NVLAP accredited calibration included.
? The temperature range may be limited by the reference probe connected to the readout. The Built-In Reference Thermometer Readout Accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterisation errors.
? Measurement accuracy specifications apply within the operating range and assume 4-wires for PRTs. With 3-wire RTDs add 0.05 ohms to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
For more information on Fluke 9143-X-P Process Field Metrology Well talk to Cuthbertson Laird Group