Sign In
The Fluke 9150 thermocouple furnace can be used for calibrating thermocouple and RTD temperature probes. Calibrations may be done over a range of 150°C to 1200°C (302°F to 2192°F). Temperature display of the 9150 is 0.1 degrees below 1000° and 1 degrees above 1000°.
The temperature is accurately controlled by Fluke’s hybrid analog/digital controller. The controller uses a Type S thermocouple as a sensor and controls the well temperature with a solid state relay (triac) driven heater.
The LED front panel continuously shows the current well temperature. The temperature may be easily set with the control buttons to any desired temperature within the specified range. The calibrator’s multiple fault protection de-vices insure user and instrument safety and protection.
With interchangeable temperature blocks, you can check thermocouples as small as 1/16 of an inch in diameter. The Fluke 9150 works with 115 or 230 VAC power.
FEATURES
Low-cost thermocouple furnace.
NIST-traceable calibration included.
RS-232 port standard
Fluke 9150 Specifications
General Specifications
Temperature range
150°C to 1200°C
(302°F to 2192°F).
Display resolution
0.1° to 999.9°
1° above 1000°.
Stability
+0.5°C.
Display accuracy
+5°C.
Well diameter
1.25 in (32 mm).
Well depth
140 mm (5.5 in); (101 mm [4 in] removable insert plus 38 mm [1.5 in] insulator).
Heating time
35 minutes to 1200°C.
Cooling time
140 minutes with block.
Well-to-well uniformity
+2.5°C (Insert “C” at 1200°C).
Stabilisation
20 minutes.
Power
115 V AC (+10%), 10.5 A or 230 V AC (+10%), 5.2 A, switchable, 50/60 Hz, 1200 W.
Size (H x W x D)
315 x 208 x 315 mm
(12.4 x 8.2 x 12.4 in).
Weight
13 kg (28 lb).
NIST-traceable calibration
Data at 150°C, 300°C, 450°C, 600°C, 800°C, 1000°C, and 1200°C.
For more information on Fluke 9150 Thermocouple Furnace Model talk to Cuthbertson Laird Group