NB: ‘X’ in the model number to be replaced with A, B, C, D, E, and F as appropriate for the desired insert.
The Fluke 9190A ultra-cool field metrology well from Fluke Calibration provides excellent stability and uniformity and plenty of well depth for calibrating platinum resistance thermometers (PRTs), thermocouples and other industrial thermometers.
The Fluke 9190A Ultra-Cool Field Metrology Well is the most accurate and stable, cold temperature dry-block on the market. It’s ideal for applications that demand strict quality control and regulatory process compliance. These applications include on-location validation and calibration of RTDs, thermocouples, thermometers, and sensors used with process control equipment such as medical freezers, laboratory refrigerators, cold rooms, blood banks, sterilisers (autoclaves), and freeze dryers.
The 9190A is small and lightweight, making it easy to transport. And since it does not use heat transfer fluids, cleanrooms stay clean. The 9190A cooling and heating times are faster than a calibration bath - that means calibration work gets done more quickly.
The Calibrator has two models: the 9190A and the 9190A-P. The 9190A-P is a “Process” version of the 9190A that combines the heat source with a built-in thermometer input panel.
FEATURES
Dual heater zone: Active heater zone control compensates for differential temperatures between zones.
All controls and inputs on the front panel.
Two-handed carry; built-in handles on front and back. All controls and inputs on the front panel.
Define up to eight set-point temperatures; automatically ramp to and “soak” at each set-point.
Stability indicator: Indicates when the block temperature is stable and within limits defined by the user.
Fast time to temperature and small footprint: Uses compact, efficient free piston stirling cooler technology (FPSC).
Remote operation: USB and RS-232 communications interfaces support remote operation from a computer.
4-20 mA connectors: Power a 4-20 mA transmitter from the drywell. Supplies 24 V dc loop power (process version).
4-wire PRT/RTD input: Reads a 4-wire, 3-wire, or 2-wire PRT/RTD. Accuracy +0.02 °C (process version).
Reference thermometer input automatically reads Fluke Calibration PRTs terminated with a “smart” connector.
Thermocouple input: Accepts common thermocouples with mini-jack termination. Types: J, K, T, E, R, S, M, L, U, N, C.
Fluke 9190A Product Video
Fluke 9190A Specifications
Temperature range at 23 °C
-95 °C to 140 °C (-139 °F to 284 °F).
Display accuracy
+ 0.2 °C full range.
Accuracy with external reference3
One year.
Stability
+ 0.015 °C full range.
Axial uniformity at 40 mm (1.6 in)
+ 0.05 °C full range.
Radial gradient
+0.01 °C full range.
Loading effect
(with a 6.35 mm reference probe and three 6.35 mm probes)
+ 0.006 °C full range
(versus display with one 6.35 mm probe)
+ 0.25 °C at -95 °C
+ 0.10 °C at 140 °C.
Operating conditions
0 °C to 35 °C, 0 % to 90 %
RH (non-condensing) < 2000 m altitude.
Environmental conditions for all specifications except temperature range
13 °C to 33 °C.
Immersion (well) depth
160 mm (6.3 in).
Well diameter
30 mm (1.18 in).
Heating time1
-95 °C to 140 °C: 40 min.
Cooling time1
23 °C to -90 °C: 80 min
23 °C to -95 °C: 90 min
140 °C to 23 °C: 60 min.
Stabilisation time2
15 min.
Resolution
0.01 °.
Display
LCD, °C or °F user selectable.
Power requirements
100 V to 115 V (+ 10 %) 50/60 Hz, 575 W
200 V to 230 V (+ 10 %) 50/60 Hz, 575 W.
System fuse ratings
115 V: 6.3 A T 250 V
230 V: 3.15 A T 250 V.
Computer interface
RS-232, USB Serial, and 9930 interface-it temperature calibration software included.
Safety
IEC 61010-1, Installation Category II, Pollution degree 2.
Electromagnetic environment
IEC 61326-1: Basic.
Refrigerants
R32 (Difluoromethane)
< 20 g, ASHRAE safety group A2L
R704 (Helium)
< 20 g, ASHRAE safety group A1.
Dimensions (H x W x D)
380 mm x 205 mm x 480 mm (14.9 in x 8.0 in x 18.8 in).
Weight
16 kg (35 lb).
Fluke 9190A warranty
One year.
1For ambient temperature of 23 °C.
2Time from when the SETPOINT is reached to when the unit is with in stability specification.
3The temperature range may be limited by the reference probe connected to the readout.
The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterisation errors.
For more information on Fluke 9190A-X Field Metrology Well talk to Cuthbertson Laird Group
Enquire Now
List your company on FindTheNeedle.