The Megger Baker MTR105 is a dedicated Static Motor Tester with Megger’s tried and trusted suite of insulation resistance tests (IR), plus all the great traditional features and reliability of Megger’s testers. It is an OFF LINE motor tester that can be used on small to medium sized rotating machines, typically up to 500 hp.
The MTR105 takes the test abilities of Megger’s proven IR test instruments adding DLRO four wire Kelvin low resistance test, inductance and capacitance tests to provide a versatile motor tester, all packaged in a robust hand held instrument, which up to now has simply not been available.
Additionally the MTR105 incorporates temperature measurement and compensation (for IR tests), motor direction of rotation plus supply phase rotation tests.
The MTR105 is powered by six AA alkaline primary cells. As an optional accessory, rechargeable NiMH AA cells can be used with an external power supply and adaptor to charge the batteries whilst in the instrument. Lithium primary (not Lithium Ion) cells can also be used.
FEATURES
Full colour graphic display.
Rotary dial control and full graphic display - simple and easy to use.
3 Phase insulation resistance.
Temperature correction for insulation resistance.
Guard terminal to eliminate any surface leakage current.
DLRO four wire Kelvin.
Continuity and diode test.
Motor direction of rotation.
Capacitance and inductance.
CAT III 600 V up to 3000 m.
Environmental protection to IP54, including the battery and fuse compartments.
Stores test results for up to 256 motors, which can be downloaded to a USB drive.
Rechargeable batteries with mains charger kit option.
Megger Baker MTR105 Static Motor Tester Video
Megger Baker MTR105 Specifications
Insulation Resistance
Accuracy
Volts Accuracy 50 V 10 GΩ +2% +2 digits +4.0% per GΩ
100 V 20 GΩ +2% +2 digits +2.0% per GΩ
250 V 50 GΩ +2% +2 digits +0.8% per GΩ
500 V 100 GΩ +2% +2 digits +0.4% per GΩ
1000 V 200 GΩ +2% +2 digits +0.2% per GΩ
Operational Uncertainty: IEC61557-2.
Polarisation Index (PI)
10 minute / 1 minute ratio.
Dielectric absorption ratio (DAR)
User configurable 15 s or 30 s t1 start time with t2 fixed at 60 s.
Guard terminal performance
<5% error at 500 kΩ parallel circuit resistance with 100 MΩ load.
Resolution
0.1 kΩ.
Short circuit / charge current
2 mA +0% -50% (IEC61557-2).
Terminal voltage accuracy
-0% +2% +2 V.
Test current
1 mA at min. pass value of insulation to a max. of 2 mA max.
Operation range
0.10 MΩ to 1.0 GΩ (IEC61557-2).
Leakage current display
0.1 µA resolution 10% (+3 digits).
Voltage display
+3% +2 digits +0.5% of rated voltage.
NOTE: Above specifications only apply when high quality silicone leads are being used - as supplied with the instrument.
Continuity
Measurement
0.01 Ω to 1 MΩ (0 to 1000 kΩ analogue scale).
Accuracy
+3 % +2 digits (0 to 99.9 Ω ) +5% + 2 digits (100 Ω - 500 kΩ).
Service error
IEC61557-4.
Test current
200 mA (-0 mA +20 mA) (0.01 Ω - 4 Ω).
Polarity
Single polarity (Default) Dual polarity (configurable on setup).
Lead resistance
Null up to 9 Ω.
Capacitance
Range
0.1 nF - 1 mF accuracy +5.0% +2 digits (1 nF - 10 µF).
Accuracy (1 nF - 10 µF):
+5.0% +2 digits.
Voltmeter
Range
dc: 0 - 1000 Vac: 10 mV - 1000 V TRMS sinusoidal (15 Hz - 400 Hz).
Accuracy
dc: +2% +2 digits (0 - 1000 V)
ac: +2% +2 digits (10 mV - 1000 V TRMS)
Operational Uncertainty: IEC61557-1.
Frequency range
15 - 400 Hz (0 - 1000 V).
Frequency resolution
0.1 Hz.
Frequency accuracy
+0.5% (+1 digit).
Diode test accuracy
+2% +2 digits 0.01 V to 3.00 V.
Display range
0.00 V to 3.00 V.
Temperature Measurement & Compensation
Thermocouple
Type T (Type K and Type J).
Thermocouple range
-20 °C to 200 °C (-4 °F - 392 °F).
Instrument range
-20 °C to 1000 °C (-4 °F - 1832 °F).
Instrument resolution
0.1 °C (0.18 °F).
Instrument accuracy
+1.0 °C +20 digits (+1.8 °F).
Low Resistance
Test current
200 mA dc.
Range
1 mΩ to 10 Ω.
Resolution
0.01 mΩ.
Accuracy
+(0.25% rdg. +10 digits).
Result Storage
Storage capacity
256 motor results (date / time stamped).
Data download
USB Type A (USB Mass Storage Device).
Power
Battery
6 x AA cells
LR6 1.5 V Alkaline
FR6 1.5 V Lithium (LiFeS2)
HR6 1.2V NiMH (rechargeable option).
Battery life
10 motors per day (complete suite of tests at 100 V into 100 MΩ)
IEC61557-2 - test cycle, 1200 insulation tests with duty cycle of 5 sec
testing on 25 sec standby @ 500 V into 0.5 MΩ
IEC61557-4 test cycle, 1200 continuity tests with duty cycle of 5 sec testing
on 25 sec standby on 1 Ω resistance.
Battery charging
Mains battery charger kit or 12 - 15 V dc vehicle charger.
Safety protection
IEC61010 CAT III 600 V.
EMC
Industrial IEC61326.
Temperature coefficient
<0.1% per °C up to 1 GΩ.
Environment
Operating temperature range
-10 °C to 50 °C (14 °F to 122 °F).
Storage temperature range
-25 °C to 50 °C (-13 °F to 122 °F).
Humidity
90% RH at 40 °C (104 °F) max.
IP rating
IP 54.
Physical
Display
Full LCD colour screen with user configurable back-light.
Dimensions
228 x 105 x 75 mm (8.98 x 4.1 x 2.95 in).
Weight
0.93 kg (2.05 lbs).
Fuse
x2 500 mA (FF) 1000 V 32 x 6 mm ceramic fuse, high break capacity HBC
Megger Baker MTR105 warranty
One-year.